U-50 Series_Máy đo nước đa chỉ tiêu hiện trường
Model: U-51 / U-52 / U-53 / U-53G /
U-54 / U-54G
Hãng sản xuất: HORIBA - Nhật Bản
Xuất xứ: Châu Á
Tham khảo tại: http://www.horiba.com/process-environmental/products/water-quality-measurement/details/u-50-multiparameter-water-quality-checker-368/
1. Tính năng kỹ thuật:
-
5
languages display (English, Japanese,
Russian, Spanish, Portuguese)
-
Multiple
sensors allow for the measurement of 11 paramenters simultaneously.
-
(pH,
pH (mV), ORP, DO, COND, Salinity, TDS, Seawater, Specific Gravity, Temperature,
Turbidity, Water depth)
-
Turbidity
sensor U-53 conforms to US EPA method 180.1 Precision has been improved over conventional
instruments. The Model U-53 high precision field replaceable sensor with wiper
has a resolution of 0.01 NTU.
-
Turbidity
sensor of U-54 conforms to EN ISO 7027. The model U-54 has a resolution of 0.01
NTU.
-
Improved
stability of the dissolved oxygen sensor has been achieved with a new 3
electrode design for fast response and polarographic sensor for ease of
maintenance.
-
pH
and ORP electrodes can be replaced individually to reduce replacement costs.
2. Thông số kỹ thuật:
U-50 series Feature comparison:
|
U-51
|
U-52
|
U-52G
|
U-53
|
U-53G
|
U-54
|
U-54G
|
|
|
pH
|
○
|
○
|
○
|
○
|
○
|
○
|
○
|
|
Dissolved
Oxygen
|
○
|
○
|
○
|
○
|
○
|
○
|
○
|
|
Conductivity
|
○
|
○
|
○
|
○
|
○
|
○
|
○
|
|
Salinity
|
○
|
○
|
○
|
○
|
○
|
○
|
○
|
|
TDS(Total
Dissolved Solid)
|
○
|
○
|
○
|
○
|
○
|
○
|
○
|
|
Seawater
specific gravity
|
○
|
○
|
○
|
○
|
○
|
○
|
○
|
|
Temperature
|
○
|
○
|
○
|
○
|
○
|
○
|
○
|
|
Turbidity
(LED)
|
—
|
○
|
○
|
—
|
—
|
○
|
○
|
|
Turbidity
(Tangsten lamp)
|
—
|
—
|
—
|
○
|
○
|
—
|
—
|
|
Water
depth
|
—
|
—
|
○
|
○
|
○
|
○*
|
○*
|
|
ORP
(Oxidation Reduction Potential)
|
○
|
○
|
○
|
○
|
○
|
○
|
○
|
|
GPS
|
—
|
—
|
○
|
—
|
○
|
—
|
○
|
Note: *U-54/G(2m) don't have the feature of water
depth.
Sensor Probe:
|
U-
51 |
U-
52 |
U-
52G |
U-
53 |
U-
53G |
U-54
|
U-54G
|
|
|
Max.items
displaying simultaneously
|
10
|
11
|
11
|
11
|
11
|
11
|
11
|
|
Measurement
temperature
|
-10
to 55°C
|
||||||
|
Max.
sensor diameter
|
Approx.
96mm
|
||||||
|
Probe
length
|
Approx.
340mm
|
||||||
|
Cable
length
|
Standard:
2m, Option: 10m,30 m
|
||||||
|
Mass
|
Approx.
1800g
(Approx. 3.97 Ibs) |
||||||
|
Automatic
calibration
(use pH4) |
Applying
to all models
|
||||||
|
Turbidity
wiper
|
–
|
–
|
–
|
○
|
○
|
–
|
–
|
|
Measurement
depth
|
Max.
30 m
|
||||||
|
Material
contact with liquid
|
PPS,
glass, SUS316L, SUS304, FKM, PEEK, Q,Titaniumu,
FEP membrane, POM |
||||||
|
Water
resistance
|
JIS
protection livel 8
|
||||||
Monitor:
|
U-51
|
U-52
|
U-52G
|
U-53
|
U-53G
|
U-
54 |
U-
54G |
|
|
Outer
dimensions
|
115(W)
X 66(D) X 283(H) mm
|
||||||
|
Mass
|
Approx.
800 g
(Approx. 1.76 Ibs) |
||||||
|
LCD
|
320
X 240 liquid crystal display with backlight (black and white)
|
||||||
|
Data
memory
|
10,000
|
||||||
|
Communication
|
USB
|
||||||
|
Battery
|
C
batteries X 4
|
||||||
|
Water
resistance
|
JIS
protection level 7 (when sensor cable is fitted)
|
||||||
|
Battery
life
|
U-51,
U-52, U-52G:
Approx. 70 hours (without backlight) U-53, U-53G: Approx.500 measurements |
Approx.
70 hours (without backlight)
|
Approx.
70 hours (without backlight)
|
||||
|
Storage
temperature
|
-10
to 60°C
|
||||||
|
Ambient
temperature
|
-5
to 45°C
|
||||||
pH(All models in common):
·
- Two-point calibration
·
- Automatic temperature
|
Specification attribute
|
Specification value
|
|
Measurement
principle
|
Glass
electrode method
|
|
Range
|
pH
0 to 14
|
|
Resolution
|
0.01pH
|
|
Repeatability
|
±0.05pH
|
|
Accuracy
|
±0.1pH
|
Oxidation Reduction Potential (ORP):
|
Specification attribute
|
Specification value
|
|
Measurement
principle
|
Platinum
electrode method
|
|
Range
|
-2000
mV to +2000 mV
|
|
Resolution
|
1
mV
|
|
Repeatability
|
±5
mV
|
|
Accuracy
|
±15
mV
|
Dissolved Oxygen (DO, All models in common):
·
- Salinity conversion (0 to 70 PPT /
automatic)
·
- Automatic temperature compensation
|
Specification attribute
|
Specification value
|
|
Measurement
principle
|
Polarographic
method
|
|
Range
|
0
to 50.0 mg/L
|
|
Resolution
|
0.01
mg/L
|
|
Repeatability
|
±0.1
mg/L
|
|
Accuracy
|
0
to 20 mg/L: ±0.2 mg/L
20 to 50 mg/L:±0.5 mg/L |
Conductivity (COND, All models in common):
·
- Auto range
- Automatic
temperature conversion (25°C)
|
Specification attribute
|
Specification value
|
|
Measurement
principle
|
4
AC electrode method
|
|
Range
|
0
to 10 S/m (0 to 100 mS/cm)
|
|
Resolution
|
0.000
to 0.999 mS/cm:0.001
1.00 to 9.99 mS/cm: 0.01 10.0 to 99.9 mS/cm: 0.1 0.0 to 99.9 mS/m: 0.1 0.100 to 0.999 S/m: 0.001 1.00 to 9.99 S/m: 0.01 |
|
Repeatability
|
±0.05%
F.S.
|
|
Accuracy
|
*±1%
F.S.
(Median of two-point calibration) |
Salinity:
|
Specification attribute
|
Specification value
|
|
Measurement
principle
|
Conductivity
conversion
|
|
Range
|
0
to 70 PPT (permillage)
|
|
Resolution
|
0.1
PPt
|
|
Repeatability
|
±1
PPT
|
|
Accuracy
|
±3
PPT
|
Total Dissolved Solid (TDS):
·
Conversion factor setting
|
Specification attribute
|
Specification value
|
|
Measurement
principle
|
Conductivity
conversion
|
|
Range
|
0
to 100 g/L
|
|
Resolution
|
0.1%
F.S.
|
|
Repeatability
|
±2
g/L
|
|
Accuracy
|
±5
g/L
|
Seawater specific gravity:
-
Display δ1, δ0, δ15
|
Specification attribute
|
Specification value
|
|
Measurement
principle
|
Conductivity
conversion
|
|
Range
|
0
to 50 δt
|
|
Resolution
|
0.1
δt
|
|
Repeatability
|
±2
δt
|
|
Accuracy
|
±5
δt
|
Temperature:
|
Specification attribute
|
Specification value
|
|
Measurement
principle
|
Thermistor
method
|
|
Range
|
-10
to 55°C
|
|
Resolution
|
0.01°C
|
|
Repeatability
|
*±0.10°C
(at calibration point)
|
|
Accuracy
|
JIS
class B platinum thermometer sensor (±0.3+0.005 |t|)
|
Turbidity (TURB):
|
U-51
|
U-52
|
U-52G
|
U-53
|
U-53G
|
U-
54 |
U-
54G |
|
|
Measurement
principle
|
–
|
LED
light source and 30°scattering method |
Tungsten
lamp
source and 90°scattering method |
LED
light source and 90°scattering method |
LED
light source and 90°scattering method |
||
|
Range
|
–
|
0
to 800 NTU
|
0
to 1000 NTU
|
0
to 1000 NTU
0 to 0.99: 0.01 NTU 1 to 99.9: 0.1 NTU |
0
to 1000 NTU
0 to 0.99: 0.01 NTU 1 to 99.9: 0.1 NTU |
||
|
Resolution
|
–
|
0
to 99.9 NTU: 0.1,
100 to 800 NTU: 1 |
0
to 9.99 NTU: 0.01,
10 to 99.9 NTU: 0.1, 100 to 1000 NTU: 1, |
0
to 0.99 NTU: 0.01,
1 to 99.9 NTU: 0.1, 100 to 1000 NTU: 1 |
0
to 0.99 NTU: 0.01,
1 to 99.9 NTU: 0.1, 100 to 1000 NTU: 1 |
||
|
Repeatability
|
–
|
*±5%
(reading) or ±0.5 NTU whichever is greater
|
±3%
(reading) or ±0.1 NTU whichever is greater
|
±5%
(reading) or ±0.5 NTU whichever is greater
|
±5%
(reading) or ±0.5 NTU whichever is greater
|
||
|
Accuracy
|
–
|
±5%
(reading) or ±1 NTU whichever is greater
|
0
to 10 NTU: ±0.5
10 to 1000 NTU: 3% (reading) or ±1 NTU whichever is greater |
±5%
(reading) or ±1 NTU whichever is greater
|
±5%
(reading) or ±1 NTU whichever is greater
|
||
Water depth:
|
U- 51
|
U-52
|
U-52G
|
U-53
|
U-53G
|
U-54
|
U-54G
|
|
|
Measurement
principle
|
–
|
–
|
Pressure
method
|
Pressure
method,
only 10m and 30m product |
Pressure
method, only 10m and 30m product
|
||
|
Range
|
–
|
–
|
0
to 30 m
|
||||
|
Resolution
|
–
|
–
|
0.05
m
|
||||
|
Repeatability
|
–
|
–
|
±1%
F.S.
|
||||
|
Accuracy
|
–
|
–
|
±0.3
m
|
||||
GPS: 12 channel parallel (This is availalble
for U-52G, U-53G and U-54G.)
Note:
*Battery life based on continuous operation using alkaline C dry batteries when the monitor temperature is over 20℃ and the backlight OFF.
*Accuracy is measured by calibrating 4 points for turbidity and electrical conductivity and 2 points for all other measurements against standard solution.
*Repeatability is measured by the ability to reproduce the results against the standard solution (at 25℃ normal pressure condition).
Note:
*Battery life based on continuous operation using alkaline C dry batteries when the monitor temperature is over 20℃ and the backlight OFF.
*Accuracy is measured by calibrating 4 points for turbidity and electrical conductivity and 2 points for all other measurements against standard solution.
*Repeatability is measured by the ability to reproduce the results against the standard solution (at 25℃ normal pressure condition).
3. Cung cấp bao gồm:
-
U-50
Series_Máy đo nước đa chỉ tiêu hiện trường, Model: U-51 / U-52 / U-53 / U-53G /
U-54 / U-54G
-
Bộ
phụ kiện tiêu chuẩn
-
Tài
liệu hướng dẫn sử dụng tiếng Anh + tiếng Việt