1.
Overview:
pH:
-
Multi-point (1...3)
calibration with up to three buffers out of eleven pre-programmed pH buffers.
-
Reads pH with 0.01
pH resolution.
mV:
-
Features mV
calibration for accurate ORP measurements.
-
Reads potentials
with 1 mV resolution.
Conductivity
-
Use a 1 cm-1electrode
to measure from 0.1 µS/cm to 100 mS/cm.
-
Use a 0.1 cm-1
electrode to measure from 0.01 µS/cm to 10 mS/cm. (C5020 only)
-
Use a 10 cm-1electrode
to measure from 1 µS/cm to 1000 mS/cm. (C5020 only)
-
Automatically
selects correct range and frequency.
-
Selectable
reference temperature: 20° or 25°C.
-
One-point
calibration.
Dissolved oxygen:
-
Operates with a
galvanic oxygen electrode requiring no polarisation time and no zero
calibration.
-
Reads dissolved
oxygen with 0.01 mg/l or 0.1% resolution.
-
Rapid air
calibration.
Temperature:
-
Reads temperatures
with 0.1°C resolution.
-
Manual or automatic
temperature compensation.
-
Calibrates
temperature probe for quality measurements.
Inputs:
-
One common input
for pH and mV.
-
One common input
for conductivity and dissolved oxygen.
-
One input for a
Pt1000 automatic temperature probe.
Display:
-
Bright LCD
screenfor better readability.
-
A white backlight
automatically illuminates when operated on the mains.
-
Stability indicator
prompts the user when readings should be taken.
-
The interactive LCD
screen provides step by step instructions in the language of your choice
(English, Dutch, French, German).
-
Shows a GLP report
on the LCD screen.
Data-logging: (C5020 only)
-
Storage memory for
300 values including temperature.
Cabinet:
-
Robust dust and
splash-proof cabinet.
Special features:
-
Three year
warranty.
-
Mains and
rechargeable battery operation with programmable auto-matic switch-off.
-
Optional 12 V car
adaptor.
Pre-programmed standards:
-
pH buffers: 1.68,
2.00, 4.00, 4.01, 6.87, 7.00, 9.18, 9.21, 10.01, 12.00, 12.45 (at 25°C).
-
Conductivity: 1413
µS/cm, 12.88 mS/cm, 111.8 mS/cm (at 25°C).
2.
Technical Data:
pH:
-
Range:
0...14 pH
-
Resolution: 0.01 pH
-
Accuracy: 0.2% ± 1 digit
-
Calibration: 1...3 points
-
Buffers: 11 pre-programmed
-
Temperature: compensation 0...100°C
-
Slope: 80...120%
mV:
-
Range: ±1000 mV
-
Resolution: 1 mV
-
Accuracy: 0.2% ± 1 digit
-
Calibration: 1
point
rH2:
-
Range: 0...42 rH2
-
Resolution: 0.1 rH2
CONDUCTIVITY:
-
Range (cc dependent):
§ 0...100 mS/cm (C5010)
§ 0...1000 mS/cm (C5020)
-
Resolution (cc
dependent):
- 0 ...µS/cm (C5010)
- 0...0.01 µS/cm (C5020)
-
Accuracy: 1% f.s.
of range
-
Calibration: 1
point
-
Standards: 3
pre-programmed
-
Cell constant (cc):
§ 1 cm-1 ±30% (C5010)
§ 0.1/1/10 cm-1±30% (C5020)
-
Temperature
compensation: 0...100°C
-
Reference
temperature: 20° or 25°C
-
Temperature
coefficient: natural waters (EN27888)
SALINITY: (C5020 only)
-
Range: 0.0...70.0
-
Reference temperature:
15°C
TDS: (C5020 only)
-
Range: 0...100 g/l
-
Resolution: 0.1
mg/l
DISSOLVED OXYGEN:
-
Range: 0...20 mg/l (0...200%)
-
Resolution: 0.01
mg/l (0.1%)
-
Accuracy: 1% ± 1
digit
-
Calibration: 1
point
-
Temperature
compensation: 0...50°C
-
Salinity
compensation: 0...40
-
Air pressure
compensation: 800...1200 hPa
TEMPERATURE:
-
Range: 0...100°C 0...100°C
-
Resolution: 0.1°C
0.1°C
-
Accuracy: 0.5°C
0.5°C
-
Calibration: 1
point 1 point
INPUTS:
-
pH/mV: BNC, 1012
Ω BNC, 1012Ω
-
Conductivity/Dissolved
oxygen: BNC
-
Temperature: 2
banana, for Pt1000 2 banana, for Pt1000
MEMORY: (C5020 only)
-
Data sets: 300 300
DISPLAY:
-
LCD: 122x32 pixels
122x32 pixels
-
White backlight
AMBIENT CONDITIONS:
-
Temperature: 0...40°C 0...40°C
-
Humidity: 0...95%,
non condensing 0...95%, non condensing
POWER SUPPLY:
-
Mains: 100...240 VAC, 50/60 Hz 100...240 VAC,
50/60 Hz
-
Low voltage: 9...15
VDC 9...15 VDC
-
Batteries: 4x1.2 V,
NiMH 4x1.2 V, NiMH
DIMENSIONS (WxDxH): 10x20x4 cm 10x20x4
cm
WEIGHT Meter: 350 g 350 g
3.
Order Info:
|
Item
|
Code
|
Description
|
|
01
|
C5010
|
pH/conductivity/DO meter
|
|
02
|
C5020
|
pH/conductivity/DO meter
|
|
03
|
C50x0P
|
Meter kit for pH: meter + pH/ATC electrode SP10T + 2x50
ml buffers (pH 4 and 7) + 50 ml electrolyte (3M KCl) + carrying case
|
|
04
|
C50x0K
|
Meter kit for conductivity: meter + conductivity/ATC
electrode SK10T + 50 ml conductivity standard (0.01 M KCl) + carrying case
|
|
05
|
C50x0Z
|
Meter kit for oxygen: meter + dissolved oxygen
electrode SZ10T + carrying case
|
|
06
|
C50x0T
|
Meter kit complete: meter + pH/ATC electrode SP10T +
conductivity/ATC electrode SK10T + dissolved oxygen electrode SZ10T + 2x50 ml
buffers (pH 4 and 7) + 50 ml electrolyte (3M KCl) + 50 ml conductivity
standard (0.01 M KCl) + carrying case
|
|
07
|
C50x0X
|
Meter kit without electrodes: meter + 2x50 ml buffers
(pH 4 and 7) + 50 ml electrolyte (3M KCl) + 50 ml conductivity standard (0.01
M KCl) + carrying case
|
|
08
|
A4049
|
Car adaptor, 12 V (optional)
|